GL240 Series

GL240 Series

Share:


GL240 Series

GL240 Main unit specifications

Item Description
Number of analog input channels 10 channels
External input/
output *1
Input *2 Trigger or Sampling (1 channel), Logic/Pulse (4 channels)
Output *3 Alarm (4 channels)
Sampling interval 10 ms to 1 hour (10ms to 50ms: voltage only) *4, External signal
Time scale of waveform display 1sec. to 24 hour /division
Trigger,
Alarm function
Trigger action Start or stop capturing data by the trigger
Repeat action Off, On (auto rearmed)
Trigger source Start: Off, Measured signal, Alarm, External, Clock, Week or Time
Stop: Off, Measured signal, Alarm, External, Clock, Week or Time
Condition Setting Combination: OR or AND
Analog signal: Rising (High), Falling (Low), Window-in, Window-out
Logic signal: Pattern (combination of each input signal in high or low)
Pulse (number of count): Rising (High), Falling (Low), Window-in, Window-out
Alarm output Outputs a signal when alarm condition occurs in the input signal *5
Pulse input
function
Rotation count
(RPM)
Counts the number of pulses per sampling interval and converts to rpm
(rotations per minute), Number of pulses for one rotation may be set to
50, 500, 5000, 50k, 500k, 5M, 50M, 500M rpm/F.S. (rpm./Full Scale)
Accumulating
count
Accumulates the number of pulses from the start of measurement
50, 500, 5000, 50k, 500k, 5M, 50M, 500M C/F.S. (Counts/Full Scale)
Instant count Counts the number of pulses per sampling interval
50, 500, 5000, 50k, 500k, 5M, 50M, 500M C/F.S. (Counts/Full Scale)
Calculation
function
Between channels Addition, Subtraction, Multiplication, and Division for analog input
Statistical Select two calculations from Average, Peak, Maximum, Minimum, RMS
Search function Search for analog signal levels, values of logic or pulse or alarm point
in captured data
Interface to PC USB (Hi-speed), WLAN (using B-568 option)
Storage
device
Media SD memory card (Support SDHC, up to 32 GB), supports 2 slots *6
Saved contents Captured data, Setting conditions, Screen copy
Capturing mode Mode: Normal, Ring, Relay
Ring: Saves most recent data (Number of captured data: 1000 to 2000000 points) *7
Relay: Saves data to multiple files without losing data until data capturing is stopped
Replay Data Replays captured data that was saved in the GL240 (in BGD or CSV format)
Scaling (Engineering unit) function Measured value can be converted to the specified engineering unit
• Analog voltage: Converts using four reference points (gain, offset)
• Temperature: Converts using two reference points (offset)
• Pulse count: Converts using two reference points (gain)
Action during data capture • Displaying parst data (using dual display mode (Current + Past data))
• Hot-swapping the SD memory card
• Saving data in between cursors
Display Size 4.3-inch TFT color LCD (WQVGA: 480 x 272 dots)
Language English, French, German, Chinese, Korean, Russian, Spanish, Japanese
Information *8 Waveform in Y-T with digital values, Waveform only, Digital value, Digital values
and statistics values
Operating environment 0 to 45 ºC, 5 to 85 % RH (non condensed)
(When operating with battery pack 0 to 40 ºC, charging battery 15 to 35 ºC)
Power source AC adapter 100 to 240 V AC, 50/60 Hz (1 pc of adapter is attached as standard accessory)
DC power 8.5 to 24 V DC (DC drive cable (option B-514) is required)
Battery pack Mountable battery pack (battery pack (option B-517): 7.2V DC, 2900mAh)
Power consumption *9 Max. 36 VA
External dimensions (W x D x H) Approx.188 x 117 x 42 mm (Excluding projections)
Weight *10 500 g