ELSO Technologies Sdn Bhd
8E, First Floor,
Jalan Tun Dr. Awang,
Bukit Jambul,
11900 Bayan Lepas,
Penang, Malaysia.
+604-6301166
+604-6301167
+604-6301168
sales@elsotech.com

Mr Kwang (General Manager)
+6016-422 8644
 
+60164228644

Johor Branch
Lake View Apartment,
Jalan Austin Perdana 1,
Taman Austin Perdana,
81100 Johor Bahru,
Johor, Malaysia.

Chong CW (Sales Manager)
+6012-4767747
cwchong@elsotech.com

Selangor Branch
Puchong, Selangor.

Glenys Teoh (Sales Manager)
+6018-2810168
glenysteoh@elsotech.com

East Malaysia Branch
Lot 35, Lot 5277,
Lorong 9C1, Jalan Kapor,
93150 Kuching,
Sarawak, Malaysia.

Clement Lau (Sales Engineer)
+6019-8497663
clementlau@elsotech.com

ELSO Technologies Pte Ltd
Singapore Branch

1 Coleman Street
#10-06 The Adelphi
179803 Singapore.

Chew Peng Seng  (Sales Manager)
+65 9663 8176
pengseng@elsotech.com

85072A 10-GHz Split Cylinder Resonator

Previous 5 / 5 Next

85072A 10-GHz Split Cylinder Resonator  Materials Test Equipment  Keysight Technologies


Key Features & Specifications

Features

  • Measures complex permittivity and loss tangent of thin film, un-clad substrate and low loss sheet materials
  • Complies with IPC test method TM-650 2.5.5.13
  • Works with Keysight resonant cavity software
  • Innovative design is robust and easy to use

Electrical Characteristics

  • TEO11 resonant mode of closed empty cylinders: frequency = 10.03 ± 0.03 GHz
  • TEO11 resonant mode of closed empty cylinders: Q factor = ≥ 20,000 (input coupling at -50dB)

Typical Uncertainty

  • TE011 mode: real part of permittivity: ±1%
  • TE011 mode: loss tangent: ±0.0001
  • Useable higher order TE0np modes: real part of permittivity: ±1 – 2%
  • Useable higher order TE0np modes: loss tangent: ±0.0005

Sample Requirements

  • Non-magnetic (µr* = 1-j0), homogeneous and non-isotropic, with uniform thickness and flat parallel sides
  • Thickness: 0.1mm to 5mm, typically 1mm
  • Minimum length or diameter 56 mm, typically >60 mm
  • Suggested values for 1mm thick sample are: real part of permittivity <100, loss tangent <0.01

Description

The Keysight 85072A 10-GHz split cylinder resonator measures permittivity and loss tangent of thin film, un-clad substrates and low loss sheet materials according to the IPC TM-650 2.5.5.13 test method. Designed for robustness and ease of use, it features precision cylinders to ensure high Q factor and loss tangent resolution. Compatible with the Keysight 85071E-300 materials measurement software, the 85072A and can be purchased separately or as part of a complete turn-key solution.